MGA-9000
Modular Multi-Gas Analyzer
- 19" standard rack enclosure (IP20). Wall-mounted enclosure is designed for weatherproof protection (IP65/NEMA4X compliant).
- Options of Vortex Cooler, Thermoelectric (Peltier) Cooler, or Circulation Fans for indoor or outdoor applications.
- With Taiwan TS Type Approval, offers explosion-proof certification options (Zone 1: Ex px IIC T6 Gb, Zone 2: Ex pz IIC T6 GC).
- Integrated sample flow and pressure monitoring.
- Modular design for easy service and maintenance.
- User-friendly touchscreen with color display.
- Measures up to six parameters in one analyzer.
- Measuring chambers are heated for stable performance.
- Raw data (measured or external) storage for over three years via SD card and/or USB external storage device.
- Optional 4-20 mA isolated analog outputs for each measured component (including external inputs).
- Modbus/TCP protocol and OPC UA over RJ-45 Ethernet port as standard.
- Third-party analyzers could be coordinated for data saving, operation, etc.
Kinetics Product
The MGA-9000 is the most powerful and advanced analyzer which measures multiple gases such as NO, NO2, SO2, CO, CO2, and O2 simultaneously. With a modular design, by changing the analytical modules, it also could measure gases like CH4, C2H4, C3H8, Cl2, HCl, NH3, H2O, H2, H2S, etc. It is designed especially for environmental continuous emissions monitoring and process gas applications.
Technical Specification
Model
MGA-9000
Measuring principle and range
Per selected sensors of NDUV, UVRAS, UV-DOAS, NDIR, IR-Microflow, ZrO2, or Paramagnetic O2.
"TDLS Laser module" is configurable for single or dual range options.
Warm-up period
Approx. 60 min. depending on sensor configuration
Calibration functions
Automatically or manually, Multi-points Calibration
Measuring range
Single or dual ranges, configurable freely
Response time for T90
Typical < 15 s, depends on sensor configurations
Sample flow required
0.5 - 1 L/min.
Sample gas inlet pressure
1 A.T.M (abs)
Sample gas outlet pressure
Without back pressure
Sample gas temperature and characteristics
0 - 50 ℃, non-condensing; dry, oil-free, and dust-free gas
HMI (Display)
7" color touch screen
Analog signals (Options)
Isolated 4-20 mA outputs (Single or Dual outputs), Isolated 4-20 mA inputs
Digital signals
8 x Relay Outputs (Dry Contact) and 8 x DI (more on request), configurable freely for alarm, calibration, etc.
Communication interfaces
RJ-45 Ethernet (Modbus or OPC UA); Alarm notice via LINE messages or Emails
Wireless operation (Optional)
WLAN 802.11 a, b, g, n, d (2.4GHz and 5GHz); Bluetooth v2.1 or v4.0, 100 meters max.
Ambient temperature
0 - 50 ℃, 0-99% RH (no condensation)
Storage temperature
-20 - 60 ℃ (no condensation)
Enclosure cooling options
Vortex Cooler, Thermoelectric (Peltier) Cooler or Circulation Fans
Power supply
100 - 240 VAC, 50/60 Hz
Power consumption
Max. 240 VA
Sample gas connection
1/4" OD 316SS Double Ferrule Connector
Housing size and rating
19" rack mounting (4.5U) IP20 or wall-mounting type, IP65/ NEMA4X
Dimensions of enclosure (H x W x D)
19" rack: 200 x 426 x 570 mm
Wall-mounting: 600 x 500 x 250 mm
Weight
19" rack: Approx. 20 kg
Wall-mounting: Approx. 30 kg
Explosion proof (optional)
Taiwan TS Type Approval, for Ex Zone 1 (Ex px IIC T6) or Ex Zone 2 (Ex pz IIC T6)
Data storage
Data storage over three years via SD card or USB thumb drive
Regulation approval
US EPA 40CFR Part 60 and Taiwan EPA CEMS requirements compliance